184

AC Voltage

Range

50mV, 500mV, 5V, 50V, 700V 5 ranges

Accuracy

50mV, 500mV, 5V, 50V, 700V ±(1.5% + 15 digits) at 50Hz-
1kHz

AC Current*

Range

50mA, 500mA, 10A 3 ranges

Accuracy

50mA, 500mA, ±(1.5% + 15 digits) at 50Hz-1kHz
10A ±(3% + 15 digits) at 50Hz-1kHz

* Measure range: >10mA

Resistance*

Range

500Ω, 5kΩ, 50kΩ, 500kΩ, 5MΩ, 5 range

Accuracy

500Ω, 5kΩ, 50kΩ, 500kΩ ±(0.3% + 3 digits)
5MΩ ±(0.5% + 5 digits)

* Measure range: 50Ω to 5MΩ

Diode Test

Maximum forward voltage 1.5V, Open voltage 2.8V

Temperature (Thermocouple)*

Range

-50°C - + 1000°C

Resolution

0.1°C

Thermocouple

B, E, J, K, N, R, S, T

*

Specifications do not include probe accuracy. Temperature specifications

only apply to the GDS-307/310/320.

Continuity Beeper

< 15 Ω

Continuity Beeper

Functions

Auto Range, Max, Min, Hold, Trend plot