120

Max/Min

Individually sets the offset for
the upper and lower templates.
These are based on the ground
level of the source channel.

Auto

Sets the tolerance for the upper
and lower template an equal
distance from the source
channel. This is set to 0.4% by
default.

Save Template

7. To save or recall a template, press the Option

button when in the Template Edit menu.

For the Max/Min option, there are a number of
recall options:

From

Allows you to recall either an
internally saved reference
waveform (Wave 1 - Wave 2),
or to recall a reference
waveform using the file
manager.

To

Recall the waveform to R1
(upper template) or R2 (lower
template).

Recall

Recalls the selected reference
waveform to the corresponding
template.

Save options:

Save Max to R1

Saves the Max template to
reference 1.

Save Min to R2

Saves the Min template to
reference 2.

For the Auto option, there is a single save
option, there is not a recall option: