Fuse Replacement

Background

The GDS-200/300 has fuse protection for both
current ports, a 10A and a 1A fuse.

Fuse Type

1A: T1A 500V

10A: T10A 500V

Steps

1. Turn the power switch off on the GDS-200/300.

2. Remove the 4 screws that hold the back panel

in place.

3. Remove the back panel carefully. You will need

a case opening tool to pry open the case after
the screws are removed.
Note: the fan wiring is connected to the back
panel.

4. Replace the fuse that has been blown, as shown

below (bottom right-hand side).