Xl™ series – Xylem XL Series H-522 - Plus User Manual
Page 143

XL™ Series
SDI-12 Operations 14-27
Example of a "User Units Slope" command:
Command
Response
Time Values
Description
-------------------
------------------------
------
---------
---------------------------
"aXRS!"
"a0021
2 sec
1
Load sensor buffer with
the User Units Slope
coefficient
Subsequent Command
Response
--------------------------- ---------------------------
"aD0"
a+xxx.xxx
Key
Description
-------------
---------------------
xxx.xxx
User Units Slope
14.7.12 Read "User Units Slope" Command
The Read User Units Slope command reads the "User Units Slope" coefficient term. The value
is read from non-volatile EEPROM within the sensor and loaded into the sensor data buffer. It
can be viewed by issuing a "D" command.
Command
Response
Description
-------------
-------------------------
---------------------------
"aXRS!"
"atttn
Load sensor buffer with User Units Slope
Where:
a
Is the sensor address ("0" to "9", "A" to "Z", or "*").
XRS
Are upper-case ASCII characters.
ttt
Is a three digit integer (000-999) specifying the maximum time, in seconds, the
sensor will take to complete the command and have data available in its buffer.
n
Is a single digit integer (0-9) specifying the number of values that will be placed in
the data buffer.