Xl™ series – Xylem XL Series H-522 - Plus User Manual
Page 141

XL™ Series
SDI-12 Operations 14-25
Example of a "Write User Units Slope" command:
Command
Response
Time Values
Description
----------------
--------------------
------
-------
-----------------------------
"aXWS1.234!"
"a0021
2 sec
1
Write user units slope
coefficient
14.7.10 Write "User Units Slope" Command
The Write User Units Slope command loads the "User Units Slope" coefficient term. The new
value is stored in non-volatile EEPROM within the sensor. Once the new slope term is written to
the EEPROM, a copy is sent to the sensor data buffer for verification. This data can be viewed
by using the "D" command. To verify the current "User Units Slope" any other time, use the
"Read User Units Slope" command.
Command
Response
Description
-------------
-------------------------
---------------------------
"aXWSddd!"
"atttn
Write User Units Slope coefficient
Where:
a
Is the sensor address ("0" to "9", "A" to "Z", or "*").
XWS Are upper-case ASCII characters.
ddd
Is the new User Units Slope value. The input format is very flexible. Some
examples are shown below.
20.095
0.195
7.984E+10
167.824E5
005.9357E-7
500
ttt
Is a three digit integer (000-999) specifying the maximum time, in seconds, the
sensor will take to complete the command and have data available in its buffer.
n
Is a single digit integer (0-9) specifying the number of values that will be placed in
the data buffer.
14.7.11 Write "User Units Offset" Command