XL™ Series

SDI-12 Operations 14-21

Response to a "aM!" command: (when the system is set to factory defaults)

Command

Response

Time Values

Description

-------------

---------------------

------- --------

-------------------------------

"aM!"

"a0063"

6 sec

3

Initiate pressure and temperature
measurement

Subsequent Command

Response

--------------------------- ------------------------------------

"aD0!"

a+xxx.xxx+yyy.yyy+zz.z


Key

Description

Units

-------------

-------------------------

-----------------------------------

xxx.xxx

Current Stage

Feet (Default)

yyy.yyy Current Pressure

Pounds-per-Square Inch (PSI)

zz.z

Current Temperature

Centigrade (EC)

Note: The number of parameters returned and their order, can be modified by the user.

Response to a "aM1!" command:

Command

Response

Time Values

Description

-------------

---------------------

------- --------

-----------------------------

"aM1!"

"a0084"

8 sec 4

Initiate diagnostic pressure and
temperature measurement

Subsequent Command

Response

--------------------------- --------------------------------------------------------

"aD0!"

a+xxx.xxx+yy.y+XXXXXX+YYYYYY


Key

Description

Units

-------------

-------------------------

-----------------------------------

xxx.xxx

Current Pressure

Pounds-per-Square Inch (PSI)

yy.y

Current Temperature

Centigrade (EC)

XXXXXX

Current Pressure

Raw Reading

YYYYYY

Current Temperature

Raw Reading

H-350XL™ Supported Measurement Commands:

14.7.6 Initiate Verify Command

The Verify Command causes a verify sequence to be performed. The result of this command is