beautypg.com

Xl™ series – Xylem XL Series H-522 - Plus User Manual

Page 135

background image

XL™ Series

SDI-12 Operations 14-19

Response to a "aI!" command:

"a10 DAAH350XLvvS#nnnnnnVkkk"

H-350-XL implementation of optional field:

S#nnnnnnVkkk

(12 bytes total)

Where:

"nnnnnn"

is a six character sensor serial number

"kkk"

is a three digit sensor firmware revision level

14.7.4 Send Identification Command

The Send Identification command responds with sensor vendor, model, and version data. Any
measurement data in the sensor's buffer is not disturbed.

Command

Response

-------------

------------------------------------------------------------

"aI!"

"allccccccccmmmmmmvvvxx...xx"

Where:

a

Is the sensor address ("0" to "9", "A" to "Z", or "*").

I

Is an upper-case ASCII character.

ll

Is the SDI-12 version compatibility level, e.g. version 1.0 is represented as
"10".

cccccccc

Is an 8 character vendor identification to be specified by the vendor and
usually in the form of a company name or its abbreviation.

mmmmmm

Is a 6 character field specifying the sensor model number.

vvv

Is a 3 character field specifying the sensor version number.

xx...xx

Is an optional field of up to a maximum of 13 characters to be used for
serial number or other specific sensor information not relevant to operation
of the data recorder.

This manual is related to the following products: