Xl™ series – Xylem XL Series H-522 - Plus User Manual
Page 135

XL™ Series
SDI-12 Operations 14-19
Response to a "aI!" command:
"a10 DAAH350XLvvS#nnnnnnVkkk
H-350-XL implementation of optional field:
S#nnnnnnVkkk
(12 bytes total)
Where:
"nnnnnn"
is a six character sensor serial number
"kkk"
is a three digit sensor firmware revision level
14.7.4 Send Identification Command
The Send Identification command responds with sensor vendor, model, and version data. Any
measurement data in the sensor's buffer is not disturbed.
Command
Response
-------------
------------------------------------------------------------
"aI!"
"allccccccccmmmmmmvvvxx...xx
Where:
a
Is the sensor address ("0" to "9", "A" to "Z", or "*").
I
Is an upper-case ASCII character.
ll
Is the SDI-12 version compatibility level, e.g. version 1.0 is represented as
"10".
cccccccc
Is an 8 character vendor identification to be specified by the vendor and
usually in the form of a company name or its abbreviation.
mmmmmm
Is a 6 character field specifying the sensor model number.
vvv
Is a 3 character field specifying the sensor version number.
xx...xx
Is an optional field of up to a maximum of 13 characters to be used for
serial number or other specific sensor information not relevant to operation
of the data recorder.