Smooth out attachment point and take proximal contacts into account
Extra-oral polishing of proximal areas and larger surfaces
Finishing
For finishing glass-ceramics, appropriate grinding instruments are indispensable. If the wrong grinding instruments are
used, marginal chipping and local overheating my occur.
The following procedure is recommended for finishing IPS Empress CAD restorations:
– Use only suitable, fine-grained (grain size < 60 µm), ceramic-bonded grinding instruments or diamonds at a speed of up
to 20,000 rpm and little pressure.
– Overheating the glass-ceramic must be prevented.
– Smooth out attachment point of the holder and take the proximal contact points into account.
– Carry out individual shape adjustments, if required.
– Try in and adjust occlusion/articulation, if necessary.
– Extra-oral polishing of proximal areas and large surfaces prior to cementation.
Etching of the IPS Empress CAD restoration using IPS Ceramic Etching Gel
Silanating the IPS Empress CAD restoration using Monobond-S
Adhesive cementation
Conditioning the IPS Empress CAD restoration
To condition it for placement, etch the IPS Empress CAD restoration for 60 seconds using IPS Ceramic Etching Gel and
subsequently clean it under running water. Then, silanate the restoration using Monobond-S. Observe the corresponding
safety instructions.