CTOG-250 Comtech Traffic Optimization Gateway

Revision 1

Ethernet-based Remote Product Management

MN-CTOG250

7–69

Modem

Unit Name – The product (e.g., CTOG-250) is identified here.

System Contact / System Location – E-mail and telephone contact information for Comtech

EF Data Customer Support are provided here.

Enter the desired information. Click [Submit] to save.

Set Time (hh:mm:ss) – Use the HH:MM:SS time format to set the time, where HH=hour [00

to 23]; MM=minutes [00 to 59]; and SS=seconds [00 to 59]).

Set Date (dd/mm/yy) – Use the DD/MM/YY European time format to set the date, where

DD=day [01 to 31]; MM=month [01 to 12]; and YY=year [00 to 99].

Set the Time and Date, and then click [Submit] to save.

Circuit ID – Enter a Circuit ID string consisting of 4 to 24 characters, and then click [Submit]

to save.

G.703 Clock Extended Mode – Use the drop-down menu to select this operating mode as Off

or TxLock, and then click [Submit] to save.

10 MHz Internal Adjustment – Enter a value from (-)999 to (+)999 to set the adjustment for

the Internal 10 MHz High Stability Reference.

External Reference Frequency – Use the drop-down menu to select this operating mode as

Internal, External 1/2/5/10 MHz, or Internal 10 MHz plus External.

Set the 10 MHz Internal Adjustment and External Reference Frequency parameters, and

then click [Submit] to save.

Test Mode – Use the drop-down menu to select this operating mode as follows:

o Normal Mode – This mode clears any test modes or loopbacks, and returns the unit to its

operational state.

o Tx CW – This test mode forces the modulator to transmit a pure carrier (unmodulated).

o Tx Alt 1/0 – This test mode forces the modulator to transmit a carrier modulated with an

alternating 1,0,1,0 pattern, at the currently selected symbol rate. This causes two discrete

spectral lines to appear, spaced at +/- half the symbol rate, about the carrier frequency.

Use this mode to check the carrier suppression of the modulator.

Select the desired mode, and then click [Submit] to execute.