4.2.14 OUTPUT CAL FACTORS Menu

Figure 4-27: OUTPUT CAL FACTORS menu

The OUTPUT CAL FACTORS menu provides access to the output calibration parameters.
These parameters are password protected and can only be changed after the calibration
password has been entered. For three phase configurations, the PHASE keys toggle between
the three calibration screens for each phase. Refer to the calibration section in this manual for
details on performing a calibration.

The following calibration factors are available from this menu:

Entry

Description

VOLT FS

Full scale voltage output calibration factor.

V HI-FREQ

High frequency full scale voltage output calibration factor.

VOLT ZERO

Zero offset voltage calibration factor.

PHASE OFST

Phase offset calibration factor. Compensates for phase shift
caused by AC amplifier.

IMP. REAL FS

Full scale resistive output impedance calibration factor.

IMP. REACT FS

Full scale inductive output impedance calibration factor.

IMP. REAL MIN

Minimum resistive AC source output impedance. The AC source
has an output impedance greater than zero. This value
determines the minimum resistive component of the AC source
output impedance.

IMP. REACT MIN

Minimum reactive AC source output impedance. The AC source
has an output impedance greater than zero. This value
determines the minimum inductive component of the AC source
output impedance.