beautypg.com

AMETEK i Series User Manual

Page 259

background image

User Manual

i Series / iX Series

245

410Hz for 5 sec.

407Hz for 4 sec.

LOW FREQUENCY

This test will change the output frequency in the following sequence:

375Hz for 1 sec.

380Hz for 4 sec.

390Hz for 5 sec.

393Hz for 4 sec.

9.8.1.7.3 EMERGENCY TEST

From the MIL704 main menu (Figure 9-53) scroll to the EMERGENCY AC entry using the up and
down cursor keys. Press the ENTER key to select the EMERGENCY screens. The screen will
appear as shown in Figure 9-56.

Figure 9-56: Emergency Menu

The EMERGENCY SCREEN has the following tests:

1

VOLTAGE

2

FREQUENCY

The above tests can be selected by scrolling to the highlighted selection using the up and down
key and the ENTER key to start the selected test.

VOLTAGE

This test will change the output voltage in the following sequence:

108V for 1 minute.

118V for 1 minute.

115V for 1 minute.

The

key (backspace) will terminate the test at any time.

FREQUENCY

This test will change the output frequency in the following sequence:

393Hz for 1 minute.

407Hz for 1 minute.

400Hz for 1 minute.

The

key (backspace) will terminate the test at any time.