AMETEK i Series User Manual

Page 150

background image

User Manual

136

i Series / iX Series

9.1.1.5 AC TESTS

Note: Prior to test selection the standard and the group selection are required. Use the shuttle to
select the standard and the group if applicable.

9.1.1.5.1 Normal state test

Scroll to the NORMAL STATE AC entry using the up and down cursor keys. Press the ENTER
key to select the NORMAL STATE screens. The screen will appear as shown in Figure 9-3.

Figure 9-3: Normal state

The DO160 NORMAL 1 and NORMAL 2 screens have the following tests:

1

VOLT FREQ MIN

2

VOLT FREQ MAX

3

VOLT UNBALANCE

4

WAVEFORM DISTORTION

5

VOLT MODULATION

6

FREQ MODULATION

7

POWER INTERRUPT

8

VOLTAGE SURGE

9

FREQ TRANSIENT (group 1)

FREQ VARIATION (group 2 & 3)

The above tests can be selected by scrolling to the highlighted selection using the up and down
key and the ENTER key to start the selected test. For some of these tests, numeric data entry
may be required to define the test number or the modulation rate.

VOLT FREQ MIN

Standard/Group

RTCA

Group1

Group2

Group3

Voltage

1

100

104

104

104

3

101.5

105.5

105.5

105.5

Frequency

380

390

360

360

Table 9-1: Normal Voltage and Frequency minimum

Standard/Group

RTCA

Group1

Group2

Group3

Voltage

1

122

122

122

122

3

120.5

120.5

120.5

120.5

Frequency

420

410

650

800

Table 9-2: Normal Voltage and Frequency Maximum

This test will set the voltage and frequency to levels defined by Table 9-1. The test will last for 30
minutes. The test will be repeated for the Group 2 and 3 using the Voltage setting from Table 9-2
and the frequency from Table 9-1. The

key (backspace) will terminate the test at any time.