POWER INTERRUPT

This test requires a numeric entry value equal to the test number. The tests are grouped as follows:

Test numbers 1 through 15 are for all Standard and Groups. See Figure 9-6 for details of the tests.

Test numbers 16 and 17 for all equipment that does not incorporate digital circuit. Test number 16 will

drop the output to zero voltage for 50 ms. Test number 17 will drop the output to zero voltage for 200
ms. Test number 18 is used for DC equipment and will drop the output for 1 sec.

Test numbers 21 through 26 are applicable for Groups 2 and 3 only. Output frequency will be set to the

F1 value for 1 second prior to the test. The output frequency will remain set to the F2 value when the
test is completed. This will allow the user to apply sequence of power interrupts. See Figure 9-7 for
detail of the tests.

V (NOM)

% of V NOMINAL

(V MIN)

0 VOLTS

T1

T2

T3

DO160 Table 16-1: Test conditions for equipment with digital circuits.
NOTES

1:

Definitions:

T1

Power interrupt time

T2

Time it would take for the applied voltage to decay from V (nom) to zero volts.

T3

Time it would take for the applied voltage to rise from zero to V (nom) volts.

V MIN

The minimum level (expressed as a percentage of V NOMINAL) to which the applied
voltage is permitted to decay.

2:

Tolerance to T1, T2, T3 = ± 10%

3:

Test condition numbers 8 and 15 are for category Z, dc powered equipment only.

Applicable
Category:

A

A, Z

Z

A, B, Z

A, Z

Z

Test
Condition
No.

1**

2

3

4

5

6

7

8

9

10

11

12

13

14

15

T1 (ms)

2**

10

25

50

75

100

200

1000

10

25

50

75

100

200

1000

T2 (ms)

<1

20*

20

20

20

20

20

20

50*

50*

50

50

50

50

50

T3 (ms)

<1

5

5

5

5

5

5

5

20

20

20

20

20

20

20

%V Nom.
(V min)

0

50

15

10

5

0

0

0

80

50

0

15

5

0

0

*

Voltage will not reach zero in this test condition.

**

Equipment performance standards may require to repeat test n°1 with T1 varying from 5 to 200 ms by step

defined in the test equipment performance standards (step typically comprised between 5 ms and 20 ms depending
on equipment design.

Figure 9-6: Power Interrupt