72
i Series / iX Series
4.2.11.4 LIMIT SETUP screen
Figure 4-24:LIMIT SETUP menu
The limit setup screen is not a menu but only serves to inform the user of the hardware
capabilities of the AC source. The cursor can be moved to any of the fields in this screen but
none of these fields can be changed. The following information is provided on this screen:
Entry
Description
VOLTAGE
Maximum AC rms or DC voltage available in the high voltage
range.
CUR LIMIT
Maximum AC rms current limit available in the low voltage
range.
FREQ LO
Lowest possible fundamental frequency that can be
programmed.
FREQ HI
Highest possible fundamental frequency that can be
programmed.
PHASE C
Phase angle of phase C with respect to phase A in three phase
mode. If the AC source is a single phase model, this field will
shown 0°. If the AC source is a split phase model, this field will
shown 180°.