4.2.13 MEASUREMENT CAL FACTORS Menu

Figure 4-26: MEASUREMENT CAL FACTORS menu

The MEASUREMENT CAL FACTORS menu provides access to the measurement calibration
parameters. For three phase configurations, the PHASE keys toggle between the three
calibration screens for each phase. These parameters are password protected and can only be
changed after the calibration password has been entered. Refer to the calibration section in this
manual for details on performing a calibration.

The following calibration factors are available from this menu:

Entry

Description

VOLT FS

Full scale voltage measurement calibration factor.

VOLT 0

Zero offset voltage measurement calibration factor.

VOLT FS HT

Temperature compensated full scale voltage measurement
calibration factor.

VOLT 0 HT

Temperature compensated zero offset voltage measurement
calibration factor.

CURR FS

Full scale current measurement calibration factor.

CURR 0

Zero offset current measurement calibration factor.

CURR FS HT

Temperature compensated full scale current measurement
calibration factor.

CURR 0 HT

Temperature compensated zero offset current measurement
calibration factor.