AMETEK i Series User Manual
Page 306

User Manual
292
i Series / iX Series
9.9.4 Initial Setup
The
–ABD option supports both AC and DC modes of operation. The correct mode is a function
of the EUT to be tested and the operation must select the corresponding operating mode first.
This can be done from the front panel if needed or using the CIGui32 program. Use of the
CIGui32 program for all settings will be assumed for the remainder of this chapter.
Five power groups are covered in the ABD0100.1.8 directive:
Group
Description
A
115VAC / 400 Hz (CF)
B
26 VAC / 400 Hz
C
115VAC / Variable Frequency (VF), 360
– 800 Hz.
D
28VDC, Conventional DC network.
E
28VDC, No Break Power Transfer (NBPT) DC network.
The required steady state output settings for each power group must be selected based on the
type of EUT to be tested. These settings can be made from the front panel of the I/IX Series
power source or from the main CIGui32 control screen.