beautypg.com

AMETEK i Series User Manual

Page 253

background image

User Manual

i Series / iX Series

239

9.8 Option

–704: MilStd704 Tests

The MIL704 option is made up of both firmware that resides in the power source and the
CIGui32 Windows application program. The firmware covers revision D and E of the Mil-Std704
standard. The CIGui32 covers the remaining revisions A, B, C and F. This provides coverage of
all available standard revisions.

The user interface for each implementation is different however. The revision D and E tests can
be operated directly from the power source‟s front panel or through the supplied CIGui32
program. The revision A-C, F tests can only be operated through the supplied CIGui32 program.
Thus, for rev A-C and F, a Windows PC and interface is required.

Section 9.8.1 covers operation of the firmware based revision D and E tests. Section 9.8.2
covers operation of the revision A-C and F tests. Note that future updates of the CIGui32 may
include overlapping coverage for revisions D and E in the software as the CIGui32 program is
designed to support all revisions. For now, revisions D and E have no associated data files in the
CIGui32.

This manual is related to the following products: