6.6.4 Current Monitor Offset Adjustment

1. Program the 300 Vac range, zero volts, and turn the Output ON.

2. On the current limit board, monitor TP5 with respect to TP1. Adjust R23 on the input/output

board for 0.0 mVDC 0.5 mVDC.

CAUTION: Use extreme care in adjusting R23. Do not touch any connections on the
Input/Output Board Assy. Dangerous voltages exist on most exposed connections.