If the Meister curve group is selected, the screen will appear as shown in Figure 9-31. The
screen has the following parameters that are unique to the group:

Figure 9-31: IEC 61000-4-13 Meister Curve

1.

LEVEL

Sets the percentage level of the inter harmonics relative to the

fundamental. The level is fixed for the entire frequency range defined by FRANGE. To
change levels, the test state must be off and user class must be selected.

2.

FRANGE

The range is selected by rotating the shuttle. The range selection is

required to change the level. An individual range step can be run by selecting the
desired range, followed by RUN RANGE from the RUN field.

3.

STEP

The step size defines the inter harmonics sweep points. This step size is

fixed for the entire range of the sweep. The STEP size can be changed when the test
state is off.

4.

PAUSE

Sets the time in seconds for which the test will pause between ranges.

There is only one value for the entire test. The PAUSE can be set when the test is not
running. For the Meister curve test, the pause time is normally set to 0 secs. If a value
other than 0 is entered, the inter harmonics are set to 0% during the pause times.

5.

RESONANT

This field is used to report the resonant points, if any, after running the

Meister curve test. No editing is allowed in this field. When the test is completed, a
display of harmonics current versus frequency plot is available. To view the graph, select
the field and press the ENTER key.

9.3.4.3 RUNNING THE TEST

The test sequence used is a function of the EUT class. The end user needs to determine the
appropriate class of EUT. The test protocol is document in the IEC 61000-4-13 test standard. For
reference, the test flow charts are shown here. It is assumed that the end user has a copy of the
actual test standard however.