The Output Calibration Table is a summary of the output calibration procedure. The following
text is a detailed explanation of the procedure.
150 VAC Range DC Zero:
Press the PROGRAM key and select the 150 Range with the shuttle. Program the output to 0.0
volts. Go to the Output Calibration, press the PHASE key for the phase to be calibrated and
select the VOLT ZERO parameter. Close the Output Relay by pressing the OUTPUT ON/OFF
key. Monitor the DC output voltage with the external DVM. With shuttle control adjust the output
voltage for 0.0 ± 0.005 VDC.
150 VAC Range Full-scale:
Press the PROGRAM key and program 120.0 volts and 60 Hz. Go to the OUTPUT CAL screen
by choosing the PREVIOUS screen selection. Select the VOLT FS parameter and adjust the
output to 120.0 ±0.05 volts.
150 VAC Range HI Freq:
Press the PROGRAM key and program the output to 120.0 volts and 400 Hz. Go the OUTPUT
CAL screen and select the V HI FREQ parameter. Adjust the output with the shuttle for an
output of 120.0 ±0.05 volts. Repeat the 60 Hz and 400 Hz adjustments until the output is within
±0.05 volts of the programmed value.
300 VAC Range DC Zero:
Program the output to the 300 VAC Range by pressing and selecting the 300 Range with the
shuttle. Go the OUTPUT CAL screen, select the VOLT ZERO parameter and adjust the output
to 0.0 ±0.005 VDC.
300 VAC Range Volt Full-scale:
Program the output to 240.0 volts and 60 Hz. Go to the OUTPUT CAL screen and adjust VOLT
FS parameter for an output of 240.0 ±0.05 volts.
300 VAC Range Volt HI Freq:
Program the output to 240.0 volts and 400 Hz. Go to the OUTPUT CAL screen and select the V
HI FREQ parameter. Adjust the output with the shuttle for an output of 240.0 ±0.05 volts.
Repeat the 60 and 400 Hz adjustments until the output is within ±0.05 volts of the programmed
value.
300 VDC Range Volt Zero:
Press the PROGRAM key. Select the MORE option to go to the PROGRAM 2 screen. Highlight
the VOLT MODE parameter and select the DC Mode by moving the shuttle. Press the
PROGRAM key and program the 300 Range and 0.0 volts. Go to the OUTPUT CAL screen by
pressing the MENU key three times followed by pressing the ENTER key to display the OUTPUT
CAL screen. Adjust the VOLT ZERO for 0.0 ±0.005 volts DC.
300 VDC Range + Full-scale:
Program + 240.0 volts. Go to the OUTPUT CALIBRATION screen and adjust the VOLT FS
parameter for an output voltage of +240.0 ±0.05 volts DC.
300 VDC Range - Full-scale:
Program - 240.0 volts. Go the OUTPUT CALIBRATION screen and adjust the VOLT FS
parameter for an output voltage of -240.0 ±0.05 volts DC.