AMETEK i Series User Manual
Page 237

User Manual
i Series / iX Series
223
OMNI Type
1-18i, 3-18i, 1-37i, 3-37i
1-37iJ
Impedance ± 5%
(in combination with power source output impedance)
Phase
0.24 ohms + j 0.15 ohms at 50 Hz
0.24 ohms + j 0.062 ohms at 50 Hz
Neutral
0.16 ohms + j 0.10 ohms at 50 Hz
0.16 ohms + j 0.054 ohms at 50 Hz
9.6.3.3 Controls/Indicators
ON/OFF toggle switch
Flicker On/Off push button switch
Power Lamp
Bypass Lamp
Flicker Lamp
9.6.3.4 Mechanical
Dimensions
Omni 1-18i and 3-18i: Omni 1-37i, 1-37iJ and 3-37i:
3.5"H x 19"W x 22"D 5.25”H x 19”W x 22”D
89mmH x 480mmW x 560mmD 133mmH x 480mmW x
560mmD
Weight
OMNI-1-18i
OMNI-3-18i
OMNI-1-37iJ
OMNI-3-37i
31 lb. (14 kg)
37 lb. (17 kg)
55 lb. (25 kg)
55 lb. (25 kg)
Connectors
Input AC
Impedance Network
Input and Output
System Interface
IEC 320
Kulka 9-85 series
3M 3367-1000 (mates with 3M 3366-1001)
Chassis Slide
General Devices C300S-118-U/B308 BK
9.6.4 Installation and Acceptance
9.6.4.1 Unpacking
Inspect the unit for any possible shipping damage immediately upon receipt. If damage is
evident, notify the carrier. Do not return an instrument to the factory without prior approval. Do
not destroy the packing container until the unit has been inspected for damage in shipment.
9.6.4.2 Power Requirements
The OMNI 1-18i and 3-18i use either 115 or 230 VAC to power the control circuits. The user
may select the desired input voltage using the small range selector card within the IEC input