9.5.5 Service and Maintenance

9.5.5.1 Cleaning

The exterior of the instrument may be cleaned with a cloth dampened with a mild detergent and
wrung out. Disconnect all sources of power (mains, iX-power sources, etc.) before cleaning. Do
not spray water or other cleaning agents directly on the instrument.

9.5.5.2 Fuses

There is one fuse used in the Mode-iX phase controller. It is located in the mains connector. For
continued protection against fire hazard, replace fuse with same type and rating.

Mains Voltage

Fuse Rating

Fuse Type

CI Stock Number

115 v rms, nominal

0.25 amp

3AG, slow acting

270117

230 v rms, nominal

0.125 amp

3AG, slow acting

270109