beautypg.com

AMETEK i Series User Manual

Page 300

background image

User Manual

286

i Series / iX Series

Table TAC:

Test No. Section

Subject

Test Limits

Comment

File Ref.

TAC101

Three Phase Load and Current Harmonic Measurements

115V RMS, 400 Hz

ATAC101

TAC102

Steady State Limits for Voltage (Including Unbalance) and Frequency

1

Balanced Phases

108V to 118V RMS, 380 to 420 Hz

Apply for 30 mins each test

ATAC102-1

2

Unbalanced Phases

108V to 118V RMS, +/- 3V Unbalance

Apply for 30 mins each test

ATAC102-2

TAC103

Voltage Phase Difference

Phase B 116 to 124 degrees

Apply for 30 mins each test

ATAC103

TAC104

Voltage Modulation

1

115V Nominal Test

115V RMS, 3.5V Vp-v mod

Apply for 10+ mins

ATAC104-1

2

109V Nominal Test

109V RMS, 3.5V Vp-v mod

Apply for 10+ mins

ATAC104-2

3

117V Nominal Test

117V RMS, 3.5V Vp-v mod

Apply for 10+ mins

ATAC104-3

TAC105

Frequency Modulation

1

400 Hz Nominal Test

400 Hz, +/- 4 Hz mod

Apply for 10+ mins

ATAC105-1

2

395 Hz Nominal Test

395 Hz, +/- 4 Hz mod

Apply for 10+ mins

ATAC105-2

3

405 Hz Nominal Test

405 Hz, +/- 4 Hz mod

Apply for 10+ mins

ATAC105-3

TAC106

Voltage Distortion Spectrum

Requires Additional Equipment

TAC107

Total Voltage Distortion

Voltage Distortion Factor 0.08

Apply for 30 mins

ATAC107

TAC108

DC Voltage Component

1

Positive DC Offset

115V RMS + .10V DC

Apply for 30 mins

ATAC108-1

2

Negative DC Offset

115V RMS - .10V DC

Apply for 30 mins

ATAC108-2

TAC109

Normal Voltage Transients

160V RMS to 58V RMS

1

Overvoltage Transients

Peak voltage 160V RMS

Table TAC109-2 A-G

ATAC109-1

2

Undervoltage Transients

Low voltage 58V RMS

Table TAC109-2 H-N

ATAC109-2

3

Combined Transient

160V RMS to 58V RMS

Table TAC109-2 O

ATAC109-3

4

Repetitive Normal Voltage Transient Test

90V RMS to 140V RMS

Run transient for 30 mins

ATAC109-4

TAC110

Normal Frequency Transients

350 Hz to 450 Hz

1

Overfrequency Transients

High Frequency 450 Hz

Table TAC110-2 A-D

ATAC110-1

2

Underfrequency Transients

Low Frequency 350 Hz

Table TAC110-2 E-H

ATAC110-2

3

Combined Transient

350 Hz to 450 Hz

Table TAC110-2 I

ATAC110-3

TAC201

Power Interrupt

0V for 50 msec

Table TAC201-2 A-L

ATAC201

TAC301

Abnormal Steady State Limits for Voltage and Frequency

102V RMS to 124V RMS, 370 Hz to 430 Hz Apply for 30 mins

ATAC301

TAC302

Abnormal Voltage Transients

45V RMS to 180V RMS

1

Overvoltage Transients

Peak voltage 180V RMS

Table TAC302-2 A-G

ATAC302-1

2

Undervoltage Transients

Low voltage 45V RMS

Table TAC302-2 H-N

ATAC302-2

3

Combined Transient

45V RMS to 180V RMS

Table TAC302-2 O

ATAC302-3

TAC303

Abnormal Frequency Transients

320 Hz to 480 Hz

Table TAC303-2 A-E

ATAC303

TAC401

Emergency Steady State Limits for Voltage and Frequency

104V to 122 V, 360 to 440 Hz

Apply for 30 mins

ATAC401

TAC501

No Tests

N/A to AC Utilization Equipment

TAC601

Power Failure (Three Phase)

0V for 7 seconds

Table TAC601-2 A-D

ATAC601

TAC602

One Phase and Two Phase Power Failures

1

One Phase Power Failures

0V for 7 seconds

Table TAC602-2 A-F

ATAC602-1

2

Two Phase Power Failures

0V for 30 mins

Table TAC602-2 G-J

ATAC602-2

This manual is related to the following products: