3.6.2 Remote Sense Connector TB3

When selecting external sense mode, it is important that the remote sense connections are
hooked up at the EUT or at the sense point. For single-phase systems, connect Phase A to
phase A and neutral to neutral. For three-phase system configurations, connect all three phase.

NOTE: Do not reverse or swap sense connection phasing or damage to the unit may

result.

All 3001iX and 5001iX AC Sources are shipped with the sense connections wired to the output
terminals. This will prevent a voltage fault when the external sense mode is selected. On
systems consisting of multiple 3001iX or 5001iX chassis, the end user has to connect the
external sense inputs to allow the system to operate. Some system configuration do not support
Internal sense mode in which case the sense connection must always be present at TB3.

Table 3-3: Remote Sense Connector

– TB3

Pin

Description

A

Phase A sense

B

Phase B sense

C

Phase C sense

N

Neutral sense