9.1.2.3 Specific Test Limitations and Discrepancies

While every effort has been made to provide comprehensive coverage of the RTCA/DO-160
tests, some limitations apply, largely due to hardware limitations of the iX power source such as
available programming resolution or rounding. The following table lists specific implementation
issues by test number. The deviation from the standard requirement is shown in the “Actual
setting” column. The reason for the deviation is shown in the last column.

Vnom

Test

Std requirement

Actual setting

Reason

230VAC

A(NF)

No Tests Performed

Max. output voltage limitation

230VAC

A(WF)

No Tests Performed

Max. output voltage limitation

115VAC

A(CF):16.5.1.2

115 +/- .18 Vrms

115 +/- .2 Vrms

Resolution

115VAC

A(NF):16.5.1.2

115 +/- .18 Vrms

115 +/- .2 Vrms

Resolution

115VAC

A(WF):16.5.1.2

115 +/- .18 Vrms

115 +/- .2 Vrms

Resolution

115VAC

A(CF):16.5.1.2

115 +/- 1.24 Vrms

115 +/- 1.2 Vrms

Resolution

115VAC

A(NF):16.5.1.2

115 +/- 1.24 Vrms

115 +/- 1.2 Vrms

Resolution

115VAC

A(WF):16.5.1.2

115 +/- 1.24 Vrms

115 +/- 1.2 Vrms

Resolution

230VAC

A(CF):16.5.1.2

230 +/- .36 Vrms

230 +/- .4 Vrms

Resolution

230VAC

A(CF):16.5.1.2

230 +/- 2.48 Vrms

230 +/- 2.5 Vrms

Resolution

115VAC

A(CF):16.5.1.3

400Hz mod by .75Hz 400Hz mod by .8Hz

Resolution

115VAC

A(NF):16.5.1.3

360Hz mod by .75Hz 360Hz mod by .8Hz

Resolution

115VAC

A(NF):16.5.1.3

650Hz mod by .75Hz 650Hz mod by .8Hz

Resolution

115VAC

A(WF):16.5.1.3

360Hz mod by .75Hz 360Hz mod by .8Hz

Resolution

115VAC

A(WF):16.5.1.3

800Hz mod by .75Hz 800Hz mod by .8Hz

Resolution

230VAC

A(CF):16.5.1.3

400Hz mod by .75Hz 400Hz mod by .8Hz

Resolution

230VAC

A(CF):16.5.1.5.1

320V Transient

Test skipped

Requires additional equipment

230VAC

A(CF):16.5.2.3.1

360V Transient

Test skipped

Requires additional equipment

14VDC

ADC:16.6.1.1

15.15 V nom

15.2 V nom

Resolution

14VDC

BDC:16.6.1.1

15.15 V nom

15.2 V nom

Resolution

14VDC

ZDC:16.6.1.1

15.15 V nom

15.2 V nom

Resolution

28VDC

ADC:16.6.1.2

> 5 kHz ripple voltage Test skipped

Requires additional equipment

28VDC

BDC:16.6.1.2

> 5 kHz ripple voltage Test skipped

Requires additional equipment

28VDC

ZDC:16.6.1.2

> 5 kHz ripple voltage Test skipped

Requires additional equipment

14VDC

ADC:16.6.1.2

> 5 kHz ripple voltage Test skipped

Requires additional equipment

14VDC

BDC:16.6.1.2

> 5 kHz ripple voltage Test skipped

Requires additional equipment

14VDC

ZDC:16.6.1.2

> 5 kHz ripple voltage Test skipped

Requires additional equipment

14VDC

ADC:16.6.2.1

10.25 V nom

10.3 V nom

Resolution

14VDC

BDC:16.6.2.1

10.25 V nom

10.3 V nom

Resolution

14VDC

ZDC:16.6.2.1

10.25 V nom

10.3 V nom

Resolution

14VDC

ADC:16.6.2.4

23.15V in transient

23.2V in transient

Resolution