AMETEK i Series User Manual
Page 337

User Manual
i Series / iX Series
323
Number
Message String
Cause
Remedy
6
"Limit memory lost"
Hardware
configuration settings
lost.
Contact CI service department at
instructions on restoring configuration
data.
7
"System memory lost"
Memory corrupted.
Recycle power.
8
"Calibration memory
lost"
Calibration data lost.
Contact CI service department at
instructions on restoring calibration
data or recalibrate unit.
9
"Start angle must be
first sequence"
Start phase angle in
wrong place..
Start phase angles can only
programmed at the start of a transient
list. Once a transient is in progress,
phase angle cannot be changed.
10
"Illegal for DC"
Operation not
possible in DC mode.
Switch to AC or AC+DC mode.
11
"Duplicate sequence"
N/A
N/A
12
"Too many sequence"
N/A
N/A
13
"Missing list
parameter"
One or more
transient list
parameters missing.
Check programmed lists.
14
"Voltage peak error "
Peak voltage
exceeds internal bus
voltage
This error may occur when selecting
user defined wave shapes with higher
crest factors. Reduce programmed
RMS value.
15
"Slew time exceed
dwell"
Time needed to slew
to final value is less
than dwell time.
Check dwell times in transient list
settings. Increase dwell time or
change slew rate for affected
parameter.
16
"Illegal during
transient"
Operation requested
not available while
transient is running.
Wait till transient execution is
completed or abort transient
execution first.
17
"Output relay must be
closed"
Operation not
possible with open
relay
Close relay before attempting
operation. E.g. transient execution
requires output relay to be closed.
18
"Trans. duration less
then 1msec"
Dwell time below
minimum or 1 msec
Increase dwell time to at least 1
msec.
19
"Clock and sync must
be internal"
Operation not
possible with external
clock
Switch to internal sync. (Default)
20
"Input buffer full"
Too much data
received.
Break up data in smaller blocks.
21
"EOS Fault"
Hardware error
reported by EOS
option
Cycle power on EOS to reset error. If
error persist, contact Contact CI
service department at
22
"Waveform harmonics
limit"
Harmonic contents of
user defined wave
shape is too high and
could damage
amplifier output
stage.
Reduce harmonic content or reduce
fundamental frequency programmed.