AMETEK i Series User Manual
Page 245

User Manual
i Series / iX Series
231
Scroll to the OMNI OPTION entry using the up and down courser key. Press the ENTER key.
The following MENU will appear as shown in Figure 9-47
Figure 9-47: OMNI Control Screen
Scroll to the MODE entry using the up and down courser key. USE the Shuttle Knob to engage
the OMNI. The display will indicate that the OMNI mode as shown in Figure 9-48
Figure 9-48: OMNI Control Screen
Note: The output of the AC source will drop for about 50 ms when the OMNI MODE changes
state.
To Bypass the OMNI, turn the Shuttle Knob counterclockwise.
9.6.6.4 Programmed Control
The FLICKER/BYPASS mode may be remotely programmed by the power source if the power
source and OMNI are configured for this function. A system interface cable must be connected
between the OMNI and the power source to carry the control signal to the OMNI. When
configured for programmed control, the OMNI may not be controlled by the front panel
FLICKER/BYPASS switch. Refer to the power source manual for details on programming
instructions.