9.9.13 ABD Option Related Error Messages

The CIGui32 program will monitor the power source on a regular bases for any run time errors. If
an error is reported by the power source, any test step execution will be aborted. Possible errors
and error sources are:

No

Description

Potential cause

2

Current Limit Fault

EUT may require more current than the programmed current limit
setting. This is possible on constant power loads during reduced
voltage (low line) type test steps. If the power source is set to
Constant Voltage (CV) mode, the output will trip off when the
current limit is exceeded and a Current Limit fault is generated. In
Constant Current (CC) mode, the output voltage will be reduced
to reduce the current to the set limit. In this case, no error
message will be generated.

-113

Undefined Header

Generally caused by a communication problem between the PC
and the power source. If the RS232 interface is used, it may be
necessary to increase the command delay setting of the CIGui32
program. (See Interface Settings topic of CIGui32 on-line help.)

If these problems persist, try using the GPIB interface instead.
(Requires National Instruments GPIB controller).

15

Slew exceeds dwell

The power source transient system was programmed to slew to a
voltage and/or frequency value that would take longer to
accomplish than the slew time set. This error can result in single
step mode if the steps are executed out of order or one or more
steps are skipped. The voltage/frequency values may no longer
line up. Try running the test from the first step.

16

Illegal during transient

A transient step is still in progress. Try using the Abort button to
abort any test in progress. If other CIGui32 screens other than the
Main GUI screen are open, they may have initiated a transient,
which conflicts with the ABD option. Try closing all other CIGui32
windows.

Great efforts have been made to check all test sequence files provided as part of the

–ABD

option. However, due to the breath and complexity of the ABD database, an occasional error
may occur while using these files. If an error cannot be resolved, turn on the CIGui32 bus
monitor window and rerun the test that yields the error and save the bus trace to a text file using
the File, Save All menu in the bus monitor window. This file can be emailed to

[email protected]

along with a description of the error.

Furthermore, test files were developed making reasonable assumptions concerning the intent of
the Airbus ABD0100 directive. The user is free to modify, edit, revise or otherwise update any of
these tests to match his/her own requirements. In this case, we recommend you copy the
supplied ABD_Dbase subdirectory and use this copy to make modifications to. This will preserve
the original data provided.