beautypg.com

AMETEK i Series User Manual

Page 298

background image

User Manual

284

i Series / iX Series

Test No. Section

Subject

Test Limits

Comment

File Ref.

2

Undervoltage Transients

Low Voltage 65V DC

Table HDC302-2 G-L

BHDC302-2

3

Combined Transients

65V to 475V DC

Table HDC302-2 M-N

BHDC302-3

HDC401

Emergency Limits for Voltage

240V to 290V DC

Apply each test for 30 mins

BHDC401

HDC501

Starting Voltage Transients

155V to 280V DC

Table HDC501-2 A

BHDC501

HDC601

Power Failure

0V for 7 seconds

Table HDC601-2 A-D

BHDC601

HDC602

Polarity Reversal

N/A to Rev B

This manual is related to the following products: