User Manual

8

i Series / iX Series

2.1.5 DC Measurements (iX series)

Parameter

Range

Accuracy ( )

Resolution

Voltage

0

– 300 Volts

0.1 Volts

0.01 Volt

Current

0

– 40 Amps

0.01 Amps

0.001 Amp

Power

0

– 6.000 kW

0.05 kW

0.001 kW

Current and Power Accuracy specifications are times two for 10001iX and times three for 15001iX. For 10001iX and
15001iX, resolution decreases by factor of 10, ranges for current and power increases by factor of three.

2.1.6 Harmonic Measurements (iX series)

Parameter

Range

Accuracy ( )

Resolution

Frequency fundamental

16.00 - 500 Hz

0.01% + 0.01 Hz

0.01 Hz

Frequency harmonics

32.00 Hz - 19.5 kHz

2 typ.

0.5

Voltage

Fundamental

0.25V

0.01V

Harmonic 2 - 50

0.25V + 0.1% +
0.1%/kHz

0.01V

Current

Fundamental

0.05A

0.01A

Harmonic 2 - 50

0.05A + 0.1% +
0.1%/kHz

0.01A

Accuracy specifications are times three for three phase mode. Harmonics frequency range in three phase mode is
32 Hz - 6.67 kHz. Resolution decreases by factor of 10 for 10001iX and 15001iX.

2.1.7 System Specification

Parameter

Specification

External
Modulation:

0 to 10%

Synchronization
Input:

Isolated TTL input for external frequency control. Requires 5V at 5 ma
for logic high.

Trigger Output:

400 s pulse for voltage or frequency change. Isolated output that
requires a pull-up resistor, 22K , to + 5 VDC.

Non volatile
memory storage:

8 complete instrument setups and transient lists, 32 events per list.

Waveforms

Sine (i series)

Sine, square, clipped, user defined (iX series)