beautypg.com

AMETEK i Series User Manual

Page 320

background image

User Manual

306

i Series / iX Series

Control / Display

Description

single trigger input. Subsequent repeats will trigger
automatically.

Freq F1 < ta

Sets the frequency before the transient (ta) for Table 2 type
transients. Normally this value is set automatically per Appendix
A, but the operator can change it in this field.

Freq F2 > tb

Sets the frequency after the transient (tb) for Table 2 type
transients. Normally this value is set automatically per Appendix
A, but the operator can change it in this field.

Transient
permutations data
entry grid. (Table)

This data grid displays the T1, T2 and T3 values used for each permutation of
the table values from Appendix A. These values are normally loaded
automatically when selecting a table number using the Type No. drop down list.
It is possible however to edit these values and save them to disk. User defined
transients can be recalled using the File, Open menu. See Menu help below.

Graph

The graph shows each transient type on a time scale from

–20 msec to 350

msec. Note that for AC mode, only the envelope of the AC waveform is shown
for clarity. Thus, the type of display is the same for AC and DC modes. The
graph can be zoomed using the left mouse button. Double click on the graph to
unzoom.

User Data

The area at the bottom of this tab may be used to enter general information
regarding the unit under test, the test location, operator, and environmental
conditions in effect during the test. This information will be incorporated in the
test reports generated by the GUI.

Menus

The top of the ABD Transients window contains a standard Windows menu bar. Available menus
are:

Menu

Sub Menu

Description

File

New

The selected resets all data from the Transient permutations data entry grid to
the selected Type No entry. This operation may be used to undo any edits
made by the user.

Open

Opens a specific transient test file with ABD0100_1_8 extension. This menu
may be used to directly select a test file without using the Type No. drop down
list.

Save

Saves the contents of the Transient permutations data entry grid to the
selected test file. The ABD0100_1_8 file extension is automatically appended.

Save As…

Saves the contents of the Transient permutations data entry grid to a user
defined file name. A file dialog box will appear enabling selection of the
directory and file name under which to store the contents of the test sequence
data grid. The ABD0100_1_8 file extension is automatically appended.

Print

Prints a test report using MS Word format. Only data for the presently selected
test sequence is printed.
Note: To print reports, the MS Word report format selection must be made
from the Main GUI, Options menu  Report Format menu before opening the
ABD0100.1.8 test screen. If not, an error prompt will indicate that no
supported report format is selected.

Exit

Closes the ABD Transients test window.

Edit

Cut

Cuts the selected row from the Transient permutations data entry grid. The
data in the row cut is available for a Paste operation.

Copy

Copies the selected row from the Transient permutations data entry grid.
Copied rows can be pasted to the same grid at a different location.

Paste

Pastes a copied row from to the Transient permutations data entry grid.

Run

All

Run selected test

Stop

Abort test in progress

Help

Opens this help file.