192

i Series / iX Series

HARMONICS GROUP

If the Harmonics group is selected, the screen will appear as shown in Figure 9-29. The screen
has the following parameters that are unique to the group:

Figure 9-29: IEC 61000-4-13 Harmonics

1. LEVEL

Sets the percentage level of the harmonic relative to the fundamental. Each

level is associated with a harmonic number. To change levels, the test state

must be off and user class must be selected.

2. HARM NO Selecting a harmonic number using the shuttle knob will allow examination of

the LEVEL and PHASE of each harmonic and changing the level and phase

angle when USER class is selected.

3. PHASE

Sets the phase angle of the harmonics relative to the fundamental. To change

phase, the test state must be off and user class must be selected.

4. PAUSE

Sets the pause time in seconds between harmonics tests. There is only one

value for the entire test. The PAUSE value can be changed when the test is not

running.