EXFO RTU 310-310G IP Services Test Head User Manual
Page 86
Creating and Starting a Test Case
74
FTB-8500 Series and FTB-8120NGE/8130NGE
EtherSAM (Y.1564) Test Case
To create an EtherSAM (Y.1564) test case:
1. Test configuration:
1a. Select Ethernet 10/100/1000 as the
Interface Type. Interface Type is not
available with RTU-310G.
1b. Select Single Port as the Topology.
Topology is only available with RTU-310.
1c. Select the source Clock Mode.
Available with RTU-310G only. Refer to
Clock Synchronization on page 352 for
more information.
1d. Select EtherSAM (Y.1564) as the
Application Type.
1e. Select the desired EtherSAM sub test(s). All EtherSAM sub tests
are selected by default. Refer to Overview (Configuration) on
page 322 and Services (Configuration) on page 327 for more
information.
Service Configuration Test consists in sequentially testing each
service. It validates that the service is properly provisioned and
that all specific KPIs or SLA parameters are met.
Service Performance Test simultaneously validates the quality of
all the services over time.
1f. Leave the Dual Test Set check box cleared. For Dual Test Set test
case, refer to Ethernet EtherSAM (Y.1564) and RFC 2544 Dual Test
Set Test Cases on page 80.
1g. Click Next.