EXFO RTU 310-310G IP Services Test Head User Manual
Page 321
RFC 2544 Tabs
Ethernet and Fibre Channel Application
309
Frame Loss
Test Granularity (%): Select the test granularity. The test granularity is
the percentage interval between each throughput value used for the
test. For example, 10% granularity means that the test will be
performed for 100%, 90%, 80%... of the throughput value. Choices are 1
to 10% (RFC). The default setting is 10%.
Nb of trials to average: Select the number of times the test will be
generated. Choices are 1 to 50 trials. The default setting is 1 trial.
Maximum Rate
TX-to-RX/P1-to-P2/P2-to-P1/Local to Remote/Remote to Local: Select
the maximum rate for the test. The default setting is 100 % for
10/100/1000Mbps and 10Gig-E LAN, and 92.8571428571429% for
10Gig-E WAN. P1-to-P2/P2-to-P1 is only available with RTU-310.
Select the rate’s unit. Choices are %, and Mbps/Gbps. The default
setting is %.
Minimum Test Time (Seconds): Indicates the minimum time the test
requires to run at best condition. The minimum test time is calculated
and updated once the test is started.
Unit
Interface
Rate
from
to
%
10Base-T
100Base-T
1000Base-T/X
10Gig-E LAN
10Gig-E WAN
1
1
1
1
1
100
100
100
100
92.8571428571429
Mbps
10Base-T
100Base-T
1000Base-T/X
0.1
1
10
10
100
1000
Gbps
10Gig-E LAN
10Gig-E WAN
0.1
0.1
10.0
9.2857142857142865