Operation configuration command format (wp/rp), I/o terminal setting – KEYENCE SR-750 Series User Manual

Page 78

background image

78

E SR-750 UM

NW-7 (Codabar)
settings

0900

03 to 50 Maximum read length

50

0901

03 to 50 Minimum read length

4

0902

4 to 11

Quiet zone scale factor

7

0903

0,1

Send start/stop characters

0 : Disable
1 : Enable

1

0904

0,1

Send start/stop characters letter
type

0 : as lowercase
1 : as UPPERCASE

0

0905

0,1

Inspect check-digit

0 : Disable
1 : Enable

0

0906

0,1

Send check-digit

0 : Disable
1 : Enable

1

0907

0 to 6

Check-digit type

0 : Modulus 16
1 : Modulus 11
2 : Modulus 10/Weight 2
3 : Modulus 10/Weight 3
4 : Check DR
5 : Modulus 11-A
6 : Luhn

0

JAN/EAN/UPC settings

1002

4 to 11

Quiet zone scale factor

7

1003

0,1

UPC-E reading

0 : Disable
1 : Enable

1

1004

0,1

EAN/JAN 8 digits reading

0 : Disable
1 : Enable

1

1005

0,1

EAN/JAN 13 digits reading

0 : Disable
1 : Enable

1

1006

0,1

UPC-A Output

0 : Output in 13 digits
1 : Output in 12 digits

0

1007

0,1

Add "number system" to UPC-E

0 : Disable
1 : Enable

0

1008

0,1

2-digit supplemental

0 : Disable
1 : Enable

1

1009

0,1

5-digit supplemental

0 : Disable
1 : Enable

1

1010

0,1

Ignore UPC without supplemental

0 : Disable
1 : Enable

0

1011

0,1

GTIN compatible 14-digit output

0 : Disable
1 : Enable

0

CODE128

1100

001 to 128 Maximum read length

128

1101

001 to 128 Minimum read length

1

1102

4 to 11

Quiet zone scale factor

7

1103

0,1

GS1-128

0 : Disable
1 : Enable

1

Function

Command

Number

Setting

value

Explanation

Default

Operation Configuration Command Format (WP/RP)

Operation configuration command is sent in the following format:
• Configuration change

Send command

Response

Normal process

Abnormal process

• Configuration confirmation

Send command

Response

Normal process

Abnormal process

mmm : Command number Fixed to 3 bytes
nnn… : Setting value (varies according to command number)

example) When setting the trigger input measurement method to "One-shot

trigger" (one-shot signal trigger)

Configuration change

Send command

WP, 101, 1

Response (Normal process) OK, WP

Configuration confirmation Send command

RP, 101

Response (Normal process) OK, RP, 1

„

I/O terminal setting

Function

Command

Number

Setting

value

Explanation

Default

IN
terminal
setting

IN1 terminal
function

000

0 to 8

0 : Not use
1 : Trigger input
2 : Preset Input
3 : Start test mode.
5 : Clear PLC link error
8 : Tuning operation

1

IN2 terminal
function

001

0 to 8

0 : Not use
1 : Trigger input
2 : Preset Input
3 : Start test mode.
5 : Clear PLC link error
8 : Tuning operation

2

IN1 terminal
test mode
assignment

002

1 to 3

1 : Reading rate test
2 : Processing time test
3 : Position test

1

IN2 terminal
test mode
assignment

003

1 to 3

1 : Reading rate test
2 : Processing time test
3 : Position test

1

IN1 terminal
input at
power-on

006

0, 1

0 : Not use
1 : Enable

0

IN2 terminal
input at
power-on

007

0, 1

0 : Not use
1 : Enable

0

Synchronizat
ion of IN1
and TIMING
LED

008

0, 1

0 : Disable
1 : Enable

1

Synchronizat
ion of IN2
and TIMING
LED

009

0, 1

0 : Disable
1 : Enable

0

Input polarity

010

0, 1

0 : Normal open point
1 : Normal close point

0

Input pulse
width

011

0 to 2

0 : 1 ms
1 : 2 ms
2 : 10 ms

0

WP

,

mmm

,

nnn…

OK

,

WP

ER

,

WP

,

ee

ee: Error code

RP

,

mmm

OK

,

RP

,

nnn…

ER

,

RP

,

ee

ee: Error code