beautypg.com

KEYENCE SR-750 Series User Manual

Page 23

background image

23

E SR-750 UM

ISO/IEC TR 29158 (AIM DPM-1-2006)

Decode success/failure

DEC (Decode)

Description

Evaluates whether decoding is possible or not.

Calculation
formula

Criterion

Cell contrast

CC (Cell Contrast)

Description

Evaluates the difference between the average
brightness value of bright cell (ML) and
average brightness value of dark cell (MD).

Calculation
formula

CC = (ML - MD)/ML

ML : Average brightness value of bright

cell

MD : Average brightness value of dark

cell

Criterion

A : more than 0.30
B : 0.25 to 0.30
C : 0.20 to 0.25
D : 0.15 to 0.20
F : less than 0.15

Cell modulation

CM (Cell Modulation)

Description

Evaluates the variation degree in cell
brightness

Calculation
formula

[White cell]
CM = (R - GT)/(ML - GT) (When R

≥ GT)

[Black cell]
CM = (GT- R)/(GT- MD) (When R

< GT)

R

: Brightness value

GT : Binarization threshold value
ML : Average brightness value of bright cell
MD : Average brightness value of dark cell

Criterion

Reflectance margin

RM (Reflectance Margin)

Description

Evaluates the variation degree in cell
brightness with black and white of the correct
cell considered.

Calculation
formula

[White cell]
R >= MARGIN = (R-GT)/(ML-GT)
R < GT

→ MARGIN = 0

[Black cell]
R < GT

→MARGIN = (GT-R)/(GT-MD)

R >=

→ MARGIN = 0

MARGIN : Margin of each cell
R

: Brightness value

GT

: Binarization threshold value

CC

: Cell contrast

Criterion

Fixed pattern damage

FPD (Fixed Pattern Damage)

Description

Evaluates the degree of the fixed pattern
damage (area on the right) dependent on the
code type.

Calculation
formula

Criterion

Format information damage

FID

(Format Information Damage)

Description

Evaluates the format information damage
degree of QR code.

Calculation
formula

Criterion

Version information damage

VID

(Version Information Damage)

Description

Evaluates the version information damage
degree of QR code (Model 2, version 2 and
later versions).

Calculation
formula

Criterion

Axial nonuniformity

AN (Axial Nonuniformity)

Description

Evaluates the distortion degree in vertical and
horizontal size of the code.

Calculation
formula

AN = abs (Xavg - Yavg)/((Xavg + Yavg)/2)

Xavg : Average cell size in horizontal

direction

Yavg : Vertical cell size in horizontal

direction

Criterion

A : less than 0.06
B : 0.06 to 0.08
C : 0.08 to 0.10
D : 0.10 to 0.12
F : more than 0.12

Grid nonuniformity

GN (Grid Nonuniformity)

Description

Evaluates the maximum position slip of each
cell

Calculation
formula

GN = Hmax/X

Hmax : Maximum position slip amount
X

: Cell size

Criterion

A : less than 0.38
B : 0.38 to 0.50
C : 0.50 to 0.63
D : 0.63 to 0.75
F : more than 0.75

Unused error correction

UEC

(Unused Error Correction)

Description

Evaluates the ratio of error correction unused
at the time of decoding.

Calculation
formula

UEC = 1.0 - ((e + 2t)/(d - p))

e : Number of code words that cannot be

read

t : Number of error code words
d : Number of error corrected code words
p : Number of error detected code words

Criterion

A : more than 0.62
B : 0.50 to 0.62
C : 0.37 to 0.50
D : 0.25 to 0.37
F : less than 0.25

Print growth (horizontal)

PGH

(Print Growth Horizontal)

Description

Evaluates the mark cell growth in the
horizontal direction.

Calculation
formula

(D - 0.5)/0.15

D : On the horizontal clock pattern

Ratio of the number of pixels of mark
cell

Criterion

A : -0.50 to 0.50
B : -0.70 to -0.50 or 0.50 to 0.70
C : -0.85 to -0.70 or 0.70 to 0.85
D : -1.00 to -0.85 or 0.85 to 1.00
F : less than -1.00 or more than 1.00

Print growth (vertical)

PGV (Print Growth Vertical)

Description

Evaluates the mark cell growth in the vertical
direction.

Calculation
formula

(D - 0.5)/0.15

D : On the vertical clock pattern

Ratio of the number of pixels of mark
cell

Criterion

A : -0.50 to 0.50
B : -0.70 to -0.50 or 0.50 to 0.70
C : -0.85 to -0.70 or 0.70 to 0.85
D : -1.00 to -0.85 or 0.85 to 1.00
F : less than -1.00 or more than 1.00