KEYENCE SR-750 Series User Manual
Page 44
44
E SR-750 UM
• Connection request
Sets whether the reader sends connection requests for PC host output. *Only
when the interface is set to Ethernet
Default
: Disabled
Setting range : Disable or Enable
• Port (Waiting)
Sets the port number for PC host output of the SR-750 Series. *Only when the
interface is set to Ethernet
Default
: 9004
Setting range : 1024 to 655355
* 9013, 9014, 9015, 9016, 5900, 5910, and 44818 cannot be
used as they are reserved for the main unit system.
• Keep Alive
Sets the keep-alive function of PC host output. *Only when the interface is set to
Ethernet
Default
: Enable
Setting range : Enable or Disable
Data port 2
• Interface
Sets the interface to perform PLC communication.
Default
: Not used
Setting range : Not used, RS-232C, Ethernet
• Remote IP address
Sets the IP address of the remote PLC. *Only when the interface is set to
Ethernet
Default
: 0.0.0.0
Setting range : 0.0.0.0 to 255.255.255.25
• Remote port
Sets the port number of the remote PLC. *Only when the interface is set to
Ethernet
Default
: 5000
Setting range : 1024 to 65535
PLC link
• Protocol
Sets the PLC communication protocol.
Default
: MC protocol
Setting range : MC protocol, SYSWAY (*RS-232C only), KV STUDIO, None (*RS-
232C only), TCP (*Ethernet only), EtherNet/IP (*Ethernet only) or
PROFINET (only when Ethernet is selected)
• DM front address
Sets the DM front address when a PLC link is used.
Default
: 0
Setting range : MC protocol …0 to 32767
SYSWAY
…0 to 9999
KV STUDIO …0 to 65534
• Control region address
Sets the control region address when a PLC link is used.
Default
: 0
Setting range : MC protocol …0 to 7F
SYSWAY
…0 to 6143
KV STUDIO …1 to 599
• Response region address
Sets the response region address when a PLC link is used.
Default
: 0
Setting range : MC protocol …0 to 7F
SYSWAY
…0 to 6143
KV STUDIO …1 to 599
• PLC link timing input
Sets to enable timing input by PLC link.
Default
: Disabled
Setting range : Disable or Enable
• Output length
Sets the data write length by PLC link.
Default
: 64 digits
Setting range : 1 to 100 digits
• Timing/Data wait (x 100 ms)
Sets the communication interval of reading or writing to a PLC via PLC link.
Default
: 100 ms
Setting range : 0 to 9900 ms (in units of 100 ms)
• Retry duration
Sets the interval to reconnect when PLC link communication is disabled.
Default
: 5 seconds
Setting range : 1 to 30 seconds
EtherNet/IP
• Data handshake
Sets EtherNet/IP data handshake
Default
: Disable
Setting range : Disable or Enable
• Input assembly data size (send)
Sets EtherNet/IP Input assembly (send) size within the range from 40 to 1400.
Default
: 500
Setting range : 40 to 1400
• Output assembly data size (receive)
Sets EtherNet/IP Output assembly (receive) size within the range from 4 to 1400.
Default
: 500
Setting range : 4 to 1400
• Byte swapping
Sets EtherNet/IP Byte swap for the data area
Default
: Disable
Setting range : Disable or Enable
PROFINET
• Device name
Sets the device name used for PROFINET.
Default: sr-750
• Data handshake
Sets the data handshake.
Default
: Disable
Setting range : Disable or Enable
Format
• Header
Sets the header.
Default
: Not selected
Setting range : Enter the header using 5 or less ASCII or hexadecimal
characters as desired.
• Terminator
Sets the terminator.
Default
: 0D (hexadecimal display, ASCII characters [CR])
Setting range : Enter the header using 5 or less ASCII or hexadecimal
characters as desired.
• Partition mark
Sets the partition mark.
Default
: 3A (hexadecimal display, ASCII characters ":")
Setting range : Enter any single ASCII or hexadecimal character.
• Inter delimiter
Sets the inter delimiter.
Default
: 2C (hexadecimal display, ASCII characters ",")
Setting range : Enter any single ASCII or hexadecimal characters.
• Append data size
Set this option to append data size.
Default
: Disabled
Setting range : Disable or Enable
• Append checksum
Set this option to append checksum function.
Default
: Disabled
Setting range : Disable or Enable
Important
Device name rule
1. PROFINET device name length: 1 to 240 characters
2. 1 label length: 1 to 63 characters
3. Only [a to z] (alphabet lower case), [0 to 9] (numbers), [-]
(hyphen), [.] (period) can be used for a device name.
4. [-] (hyphen) cannot be used at the beginning of the label.
5. [-] (hyphen) cannot be used at the end of the label.
6. port-xyz, port-xyz-abcde cannot be the name of the first label.
a, b, c, d, e, x, y, z means [0 to 9] (numbers).
7. Device names cannot be made in the IP address format
(n.n.n.n n=0,…,999).
8. Labels cannot start with xn-.
9. The first character of labels cannot be a number.