90
DISPLAY READS X Enter ~ Load ~ 0.00
6. Place a 100lb test weight on scale.
DISPLAY READS X Enter ~ Load ~ 99.66
7. Enter
.
DISPLAY READS X 100
8. Press
1
to establish span.
DISPLAY READS X 100.00
9. Pause for motion delay.
DISPLAY READS X Cal ~ Good? ~ 100.00
10. Press
1
to accept calibration.
DISPLAY READS X Enter ~ =Stor
11. Press
1
to save calibration.
DISPLAY READS X Enter ~ =End
12. Press
1
to exit calibration.
DISPLAY READS X 100.00
13. Remove the calibration weight.
DISPLAY READS X 0.00
First Zero Calibration Method Example (355):
1. From the Weigh Mode key in
!
.
DISPLAY READS X Setup
2. Key
in
(
.
DISPLAY READS X Fast ~ Cal~First ~ Zero? ~ -0.26
3. Remove any load on the scale.
DISPLAY READS X First ~ Zero? ~ -0.42
4. Press
(
to establish zero.
DISPLAY READS X 0.00
5. Pause for motion delay.
DISPLAY READS X Enter ~ Load ~ 0.00
6. Place a 100lb test weight on scale.
DISPLAY READS X Enter ~ Load ~ 99.66
7. Enter
.
DISPLAY READS X 100
8. Press
(
to establish span.