Three separate incrementing, non-resetable audit trail parameters are used by the Model 350/355 to indicate changes to
various parameters, P60201 – OIML, P60203 – Calibration , and P60204 – Setup. An audit trail counter will increment
only once upon exiting the Setup Mode and saving changes regardless of how many settings were changed.
OIML A
UDIT
T
RAIL
Changes to any of the following parameters will increment the OIML (Euro) audit trail at P60201:
• P110 – P118 (scale setup)
• P150 (default units)
• P151 (units enable)
• P410 (Euro enable)
• P440 (rStrc enable)
• Existing Calibration
C
ALIBRATION
A
UDIT
T
RAIL
Any changes to the existing calibration will increment the Calibration (CAL) audit trail at P60203.
S
ETUP
A
UDIT
T
RAIL
Changes to any of the Setup Mode parameters will increment the setup audit trail at P60204.
V
IEWING
A
UDIT
T
RAIL
P
ARAMETERS
Audit trail parameters may be viewed at any time.
To view audit trail parameters: (Model 350)
1. Press
+
.
DISPLAY READS X Setup~Enter ~ Code!
2. Press
1
.
DISPLAY READS X -No- ~ Chgs!~P110 ~ FS ~ xx.xx
3. Enter
[60203] with the
and
2
keys.
DISPLAY READS X 60203
4. Press
to view the selected audit trail.
DISPLAY READS X Audit ~ Trail~CAL. ~ 00001
5. Press
!
again to view the Setup Audit Trail parameter.
DISPLAY READS X Audit ~ Trail~Setup. ~ 00003
6. Press
to return to the Weigh Mode.
DISPLAY READS X 0.00
To view audit trail parameters: (Model 355)
1. Press
!