Analog output setup – Holtgreven GSE-355 I.S. User Manual
Page 32
23
6
V
9. Press
1
or
(
to increase the output to 75%.
4-20A
~
75P
8
V
10.
Press
1
or
(
to increase the output to 100%.
4-20A
~
100P
10V
Analog Output Setup
Table 3: Analog Output Parameters
Parameter
Setting
Displayed
Name
Description
Type/choices (*=Default)
P171.00
AnAlg
Enable analog option.
Disbl*, Enabl
P172.00
AnPar
Parameter that analog signal corresponds
to. Displayed corresponds to gross while
viewing the gross weight and net
otherwise.
Gross*, Net, Displayed
P173.--
AnlFS
Full scale value at which P172 selection
yields an output of 10 volts. If set to 0,
uses P110 setting.
Numeric Entry:
0* to ±1,000,000
P174.--
AnOff
Offset value which yields a 0 volt output.
Numeric Entry:
0* to ±1,000,000
P175.10
AnRng
Range Value (1-10) which specifies the
max value of analog output – entered in
terms of voltage. NOTE: This does not
change the analog span. A 0-10,000 lb.
output set to 8 will stop increasing its
analog signal at 8,000 lbs.
Numeric entry:
0 to 10*
P176.01
AnRst
Reset state – Specifies analog signal level
when Model 350 and Model 355 enters
Setup Mode.
10 V (Max Output)
0 V (Min Output)
No Change*
P177.00
AType
Specifies output type: voltage or current.
0 –10 volts*,
0 -20mA,
4 -20mA
The actual output signal is calculated as follows:
If the result is greater than the Max Range Value (P175), then the analog signal is limited to the Max Range Value.
* 10
(P172: Parm Value) - (P174: Zero Offset)
(P173: Full Scale Value) – (P174: Zero Offset)
V
out
=