8. Enter

.

DISPLAY READS X 10000

9. Press

1

to establish span.

DISPLAY READS X 10000.

10. Pause for motion delay.

DISPLAY READS X Cal ~ Good? ~ 10000.

11. Press

1

to accept calibration.

DISPLAY READS X Enter ~ =Stor

12. Press

1

to save calibration.

DISPLAY READS X Enter ~ =End

13. Press

1

to exit calibration.

DISPLAY READS X 10000.

14. Remove the calibration weight.

DISPLAY READS X 0.

Last Zero Calibration With Weight Already Applied Example (355):

1. Remove any load on the scale.

DISPLAY READS X 10.

2. Press

"

to zero the scale.

DISPLAY READS X 00.

3. Apply a 10000 lb test weight to verify calibration.

DISPLAY READS X 9970.

4. Press

!

.

DISPLAY READS X Setup

5. Press

(

.

DISPLAY READS X Fast ~ Cal~First ~ Zero? ~ 9930.

6. Press

!

.

DISPLAY READS X Last ~ Zero? ~9930.

7. Press

(

to use last zero.

DISPLAY READS X Enter ~ Load? ~ 9970.

8. Enter

.

DISPLAY READS X 10000