Holtgreven GSE-355 I.S. User Manual
Page 101

92
8. Enter
.
DISPLAY READS X 10000
9. Press
1
to establish span.
DISPLAY READS X 10000.
10. Pause for motion delay.
DISPLAY READS X Cal ~ Good? ~ 10000.
11. Press
1
to accept calibration.
DISPLAY READS X Enter ~ =Stor
12. Press
1
to save calibration.
DISPLAY READS X Enter ~ =End
13. Press
1
to exit calibration.
DISPLAY READS X 10000.
14. Remove the calibration weight.
DISPLAY READS X 0.
Last Zero Calibration With Weight Already Applied Example (355):
1. Remove any load on the scale.
DISPLAY READS X 10.
2. Press
"
to zero the scale.
DISPLAY READS X 00.
3. Apply a 10000 lb test weight to verify calibration.
DISPLAY READS X 9970.
4. Press
!
.
DISPLAY READS X Setup
5. Press
(
.
DISPLAY READS X Fast ~ Cal~First ~ Zero? ~ 9930.
6. Press
!
.
DISPLAY READS X Last ~ Zero? ~9930.
7. Press
(
to use last zero.
DISPLAY READS X Enter ~ Load? ~ 9970.
8. Enter
.
DISPLAY READS X 10000