Holtgreven GSE-355 I.S. User Manual
Page 103

94
DISPLAY READS X 2500.
9. Pause for motion delay.
DISPLAY READS X Cal ~ Good? ~ 2500.
10. Press
1
to accept calibration.
DISPLAY READS X Enter ~ =Stor
11. Press
1
to save calibration.
DISPLAY READS X Enter ~ =End
12. Press
1
to exit calibration.
DISPLAY READS X 5055.
13. Remove the calibration weight.
DISPLAY READS X 00.
False Zero Calibration Without Removing Existing Load Example (355):
1. Press
!
.
DISPLAY READS X Setup
2. Press
(
.
DISPLAY READS X Fast ~ Cal~First ~ Zero? ~ 5075.
3. Press
!
!
.
DISPLAY READS X False ~ Zero? ~5075.
4. Press
(
to establish false (temporary) zero.
DISPLAY READS X Units ~ =lb
5. Pause to display calibration units.
DISPLAY READS X Enter ~ Load? ~ 00.
6. Place a 2500lb test weight on scale.
DISPLAY READS X Enter ~ Load? ~ 2510.
7. Enter
.
DISPLAY READS X 2500
8. Press
(
to establish span.
DISPLAY READS X 2500.
9. Pause for motion delay.
DISPLAY READS X Cal ~ Good? ~ 2500.
10. Press
(
to accept calibration.
DISPLAY READS X Enter ~ =Stor