Holtgreven GSE-355 I.S. User Manual

Page 103

background image

94

DISPLAY READS X 2500.

9. Pause for motion delay.

DISPLAY READS X Cal ~ Good? ~ 2500.

10. Press

1

to accept calibration.

DISPLAY READS X Enter ~ =Stor

11. Press

1

to save calibration.

DISPLAY READS X Enter ~ =End

12. Press

1

to exit calibration.

DISPLAY READS X 5055.

13. Remove the calibration weight.

DISPLAY READS X 00.

False Zero Calibration Without Removing Existing Load Example (355):

1. Press

!

.

DISPLAY READS X Setup

2. Press

(

.

DISPLAY READS X Fast ~ Cal~First ~ Zero? ~ 5075.

3. Press

!

!

.

DISPLAY READS X False ~ Zero? ~5075.

4. Press

(

to establish false (temporary) zero.

DISPLAY READS X Units ~ =lb

5. Pause to display calibration units.

DISPLAY READS X Enter ~ Load? ~ 00.

6. Place a 2500lb test weight on scale.

DISPLAY READS X Enter ~ Load? ~ 2510.

7. Enter

.

DISPLAY READS X 2500

8. Press

(

to establish span.

DISPLAY READS X 2500.

9. Pause for motion delay.

DISPLAY READS X Cal ~ Good? ~ 2500.

10. Press

(

to accept calibration.

DISPLAY READS X Enter ~ =Stor