Holtgreven GSE-355 I.S. User Manual
Page 100

91
DISPLAY READS X 100.00
9. Pause for motion delay.
DISPLAY READS X Cal ~ Good? ~ 100.00
10. Press
(
to accept calibration.
DISPLAY READS X Enter ~ =Stor
11. Press
(
to save calibration.
DISPLAY READS X Enter ~ =End
12. Press
(
to exit calibration.
DISPLAY READS X 100.00
13. Remove the calibration weight.
DISPLAY READS X 0.00
L
AST
Z
ERO
The Last Zero procedure allows recalibration of the weighing device using an existing test load. This is especially
beneficial when checking high capacity applications such as tank weighing to minimize the task of placing and
removing test weights.
Establish gross zero before entering setup or calibration!
Last Zero Calibration With Weight Already Applied Example (350):
1. Remove any load on the scale.
DISPLAY READS X 10.
2. Press
to zero the scale.
DISPLAY READS X 00.
3. Apply a 10000 lb test weight to verify calibration.
DISPLAY READS X 9970.
4. Press
+
.
DISPLAY READS X Setup
5. Press
О
О
2
О
1
.
DISPLAY READS X Fast ~ Cal~First ~ Zero? ~ 9930.
6. Press
.
DISPLAY READS X Last ~ Zero? ~9930.
7. Press
1
to use last zero.
DISPLAY READS X Enter ~ Load? ~ 9970.