This chapter contains error messages and information parameters, as well as information on setup parameter selections
and A/D Calibration.

Error Messages

The Model 350/355 utilizes the following types of error messages: Operational Errors, Setup Mode Errors, Hardware
Errors
, Calibration Errors, Communication Errors, and Miscellaneous Errors.

O

PERATIONAL

E

RRORS

Code02

Under Load. Input signal is less than negative full scale. Check load cell wiring.
Verify correct capacity selection at P110.


Code03

Over Load. Input signal is greater than positive full scale. Use same checks as
“under load” above.


Funct ~ Disbl

Attempted to perform a function disabled in the Setup Mode.

Code 04

The digits on the display have exceed the six digit display capacity.


Code 05

Zero attempted beyond that allowed by P118.


Code 08

Input signal greatly exceeds the valid range. Check the load cell connection.


Tare ~ Error

Negative tare attempted when disabled (P440 enabled).


Tare ~ GT FS

Tare value greater than full scale capacity.


Delay

Indicates that a motion delay is in effect (zero, tare, etc.).


Delay ~ Abort

Acknowledges that a motion delayed function was aborted.

Print ~ Abort

Acknowledges that a motion delayed print request was aborted.

Add ~ Load!

If displayed after performing a count sample, this message indicates that a larger s

ample size is required.


Out of ~ Range

Attempted to enter a value beyond the allowable range.


SPtxx ~ Error

A conflict occurred with a setpoint value entry (example: target entry is less than
preact). The digits ‘xx’ represent the last two digits of the setpoint parameter
in error
(example: SPt 5 ~ Error indicates a conflict at P5105, preact 1).


Need APS

A setpoint start is initiated and the setpoint is based on quantity and no piece
weight has been established (start will not occur).