Holtgreven GSE-355 I.S. User Manual
Page 106

97
5. Place a 100lb test weight on scale.
DISPLAY READS X Code03
6. Press
.
DISPLAY READS X First ~ Zero? ~ -0.26
7. Remove any load on the scale.
DISPLAY READS X First ~ Zero? ~ -0.42
8. Press
1
to establish zero.
DISPLAY READS X 0.00
9. Pause for motion delay.
DISPLAY READS X Enter ~ Load ~ 0.00
10. Place a 100lb test weight on scale.
DISPLAY READS X Enter ~ Load ~ xx.xx
11. Enter 100 with the
and
2
keys.
DISPLAY READS X 100
12. Press
1
to establish span.
DISPLAY READS X 100.00
13. Pause for motion delay.
DISPLAY READS X Cal ~ Good? ~ 100.00
14. Press
1
to accept calibration.
DISPLAY READS X Enter ~ =Stor
15. Press
1
to save calibration.
DISPLAY READS X Enter ~ =End
16. Press
1
to exit calibration.
DISPLAY READS X 100.00
17. Remove the calibration weight.
DISPLAY READS X 0.00
Reset Calibration Gain Factors Example (355):
1. Press
!
.
DISPLAY READS X Setup
2. Press
(
.
DISPLAY READS X Fast ~ Cal~First ~ Zero? ~ xx.xx
3. Remove any load on the scale.
DISPLAY READS X First ~ Zero? ~ xx.xx