97

5. Place a 100lb test weight on scale.

DISPLAY READS X Code03

6. Press

.

DISPLAY READS X First ~ Zero? ~ -0.26

7. Remove any load on the scale.

DISPLAY READS X First ~ Zero? ~ -0.42

8. Press

1

to establish zero.

DISPLAY READS X 0.00

9. Pause for motion delay.

DISPLAY READS X Enter ~ Load ~ 0.00

10. Place a 100lb test weight on scale.

DISPLAY READS X Enter ~ Load ~ xx.xx

11. Enter 100 with the

and

2

keys.

DISPLAY READS X 100

12. Press

1

to establish span.

DISPLAY READS X 100.00

13. Pause for motion delay.

DISPLAY READS X Cal ~ Good? ~ 100.00

14. Press

1

to accept calibration.

DISPLAY READS X Enter ~ =Stor

15. Press

1

to save calibration.

DISPLAY READS X Enter ~ =End

16. Press

1

to exit calibration.

DISPLAY READS X 100.00

17. Remove the calibration weight.

DISPLAY READS X 0.00

Reset Calibration Gain Factors Example (355):

1. Press

!

.

DISPLAY READS X Setup

2. Press

(

.

DISPLAY READS X Fast ~ Cal~First ~ Zero? ~ xx.xx

3. Remove any load on the scale.

DISPLAY READS X First ~ Zero? ~ xx.xx