Holtgreven GSE-355 I.S. User Manual
Page 102

93
9. Press
(
to establish span.
DISPLAY READS X 10000.
10. Pause for motion delay.
DISPLAY READS X Cal ~ Good? ~ 10000.
11. Press
(
to accept calibration.
DISPLAY READS X Enter ~ =Stor
12. Press
(
to save calibration.
DISPLAY READS X Enter ~ =End
13. Press
(
to exit calibration.
DISPLAY READS X 10000.
14. Remove the calibration weight.
DISPLAY READS X 00.
F
ALSE
Z
ERO
False Zero calibrates the Model 350/355 without removing the current gross weight. This is particularly useful in tank
weighing applications where it may be both time consuming and costly to completely empty the tank. This operation is
achieved by establishing a false (temporary zero) zero reference. Test weights may then be added to verify calibration.
The zero reference determined during the last calibration is not affected.
False Zero Calibration Without Removing Existing Load Example (350):
1. Press
+
.
DISPLAY READS X Setup
2. Press
О
О
2
О
1
.
DISPLAY READS X Fast ~ Cal~First ~ Zero? ~ 5075.
3. Press
.
DISPLAY READS X False ~ Zero? ~5075.
4. Press
1
to establish false (temporary) zero.
DISPLAY READS X Units ~ =lb
5. Pause to display calibration units.
DISPLAY READS X Enter ~ Load? ~ 00.
6. Place a 2500lb test weight on scale.
DISPLAY READS X Enter ~ Load? ~ 2510.
7. Enter
.
DISPLAY READS X 2500
8. Press
1
to establish span.