Establishing zero – Holtgreven GSE-355 I.S. User Manual
Page 98

89
Establishing Zero
The Model 350/355 provides five methods for obtaining a zero (no load) calibration reference, First Zero, Last Zero,
False Zero, Only Zero, and Cal Reset.
To select a calibration method (350):
1. Press
+
to display the calibration prompt.
2. Press
О
О
2
О
1
to access the Calibration Mode.
3. Press
to scroll through the five selections.
4. Press
1
to establish zero.
To select a calibration method (355):
1. Press
!
to display the calibration prompt.
2. Press
(
to access the Calibration Mode.
3. Press
!
to scroll through the five selections.
4. Press
(
to establish zero.
F
IRST
Z
ERO
The most common zeroing procedure, First Zero is used to establish a new zero (no load) calibration reference before
proceeding to span the Model 350/355. Use this method for first-time calibration and complete recalibration.
First Zero Calibration Method Example (350):
1. From the Weigh Mode, press
+
.
DISPLAY READS X Setup
2. Press
О
О
2
О
1
.
DISPLAY READS X Fast ~ Cal~First ~ Zero? ~ -0.26
3. Remove any load on the scale.
DISPLAY READS X First ~ Zero? ~ -0.42
4. Press
1
to establish zero.
DISPLAY READS X 0.00
5. Pause for motion delay.
Press
or
during calibration to back up one step in the procedure.