Holtgreven GSE-355 I.S. User Manual
Page 104

95
11. Press
(
to save calibration.
DISPLAY READS X Enter ~ =End
12. Press
(
to exit calibration.
DISPLAY READS X 5055.
13. Remove the calibration weight.
DISPLAY READS X 00.
O
NLY
Z
ERO
Only Zero is used to establish a new calibration zero without affecting the span. This is useful for correcting changes to
the scale’s dead load, for example adding safety rails to a truck scale platform.
Only Zero Calibration Example (350):
1. From the Weigh Mode, press
+
.
DISPLAY READS X Setup
2. Press
О
О
2
О
1
.
DISPLAY READS X Fast ~ Cal~First ~ Zero? ~2640.
3. Press
.
DISPLAY READS X Only ~ Zero? ~ 2640.
4. Remove any load on the scale.
DISPLAY READS X Only ~ Zero? ~ 2620.
5. Press
1
to establish zero.
DISPLAY READS X 00.
6. Pause for motion delay.
DISPLAY READS X Cal ~ Good? ~ 00.
7. Press
1
to accept calibration.
DISPLAY READS X Enter ~ =Stor
8. Press
1
to save calibration.
DISPLAY READS X Enter ~ =End
9. Press
1
to exit calibration.
DISPLAY READS X 00.