Holtgreven GSE-355 I.S. User Manual
Page 31
22
6. Press
1
or
(
to set the output to 0%.
0-20A
~
0P
0.00
V
7. Press
1
or
(
to increase the output to 25%.
0-20A
~
25P
2.5
V
8. Press
1
or
(
to increase the output to 50%.
0-20A ~ 50P
5
V
9. Press
1
or
(
to increase the output to 75%.
0-20A
~
75P
7.5
V
10. Press
1
or
(
to increase the output to 100%.
0-20A
~
100P
10
V
To test the 4-20mA output mode:
Voltmeter readings are based on the use of a 500 ohm precision resistor.
Caution! Do not exceed 500 ohms.
1. Enter the Setup Mode (see Setup Mode on page 40).
Chngs
Poss!
P110.-- ~ F.S.= ~ 100.00
2. Attach the precision resistor to pin 5 and pin 6.
3. Attach the voltmeter + (red) lead to pin 6 (4-20 mA) of the analog output connector.
4. Attach the voltmeter - (black) lead to pin 5 (ISOLATED GND) of the Analog Output connector.
5. Key
in
!
(see Setup Mode on page 40).
Test ~ 4-20A
Per
P176
6. Press
1
or
(
to set the output to 0%.
4-20A
~
0P
2
V
7. Press
1
or
(
to increase the output to 25%.
4-20A
~
25P
4
V
8. Press
1
or
(
to increase the output to 50%.
4-20A
~
50P