22

6. Press

1

or

(

to set the output to 0%.

0-20A

~

0P

0.00

V

7. Press

1

or

(

to increase the output to 25%.

0-20A

~

25P

2.5

V

8. Press

1

or

(

to increase the output to 50%.

0-20A ~ 50P

5

V

9. Press

1

or

(

to increase the output to 75%.

0-20A

~

75P

7.5

V

10. Press

1

or

(

to increase the output to 100%.

0-20A

~

100P

10

V

To test the 4-20mA output mode:

Voltmeter readings are based on the use of a 500 ohm precision resistor.
Caution! Do not exceed 500 ohms.

1. Enter the Setup Mode (see Setup Mode on page 40).
Chngs

Poss!

P110.-- ~ F.S.= ~ 100.00

2. Attach the precision resistor to pin 5 and pin 6.

3. Attach the voltmeter + (red) lead to pin 6 (4-20 mA) of the analog output connector.

4. Attach the voltmeter - (black) lead to pin 5 (ISOLATED GND) of the Analog Output connector.

5. Key

in

!

(see Setup Mode on page 40).

Test ~ 4-20A

Per

P176

6. Press

1

or

(

to set the output to 0%.

4-20A

~

0P

2

V

7. Press

1

or

(

to increase the output to 25%.

4-20A

~

25P

4

V

8. Press

1

or

(

to increase the output to 50%.

4-20A

~

50P