Setup Mode Calibration
You can enter the Calibration Mode after accessing the Setup Mode to view and/or change parameter settings (see
Setup Mode on page 40).
To access the Calibration Mode when viewing any setup parameter:
1. From the Setup Mode, press
"
.
DISPLAY READS X Enter ~ =Cal!
2. Press
1
or
(
.
DISPLAY READS X First ~ Zero? ~ -0.26
Fast Calibration
Fast Calibration allows calibration of the Model 350/355 scale system without accessing the Setup Mode.
To access Fast Calibration from the Weigh Mode (350):
1. From the Weigh Mode, press
+
.
DISPLAY READS X Setup~Enter ~ Code!
2. Press
О
О
2
О
1
.
DISPLAY READS X Fast ~ Cal!~First ~ Zero? ~ -0.26
To access Fast Calibration from the Weigh Mode (355):
1. From the Weigh Mode key in
!
.
DISPLAY READS X Setup~Enter ~ Code!
2. Key
in
(
DISPLAY READS X Fast ~ Cal!~First ~ Zero? ~ -0.26
Fast Calibration can also be accessed if the following data stream is received via the comm port:
100%s54321%e
Performing Calibration
Calibration always begins by establishing a zero (no-load) reference. A complete calibration also requires establishing a
span (test load) reference. This section details various methods for obtaining zero and span references.